Siemens MAGNETOM Trio

The Siemens Trio is  a 3 Tesla MRI scanner with Tim (Total Imaging Matrix).  Its capacity enables consistent high quality imaging, the greatest efficiency for planning and processing, and will give you the confidence to diagnose the most difficult diseases quickly.  It includes Syngo Inline Technology, AutoAlign solutions, Expert-i, and MapIt.

The refurbished Siemens MAGNETOM Trio 3.0T MRI machine is a very popular MRI scanner due to its ease-of-use, signal-to-noise ratio, and its flexibility.

With Tim (Total Imaging Matrix) technology, the Siemens Trio 3.0T MRI scanner is considered one of the easier MRI scanners, and also known as a reliable, durable MRI scanner.

  • Abdominal
  • Breast
  • Cardiac
  • Colonography
  • Foot
  • Hand
  • Head
  • Knee
  • Musculoskeletal
  • Neurology
  • Oncology
  • Orthopedic
  • Shoulder
  • Spine
  • Vascular Imaging
  • Whole Body Imaging
  • Whole-Body
  • Wrist

Features

  • UltraShort bore
  • 60cm Diameter bore
  • Up to 4 body matrices
  • Open bore 70cm coil
  • Tim (Total Imaging Matrix) technology
  • 5mm to 205cm field of view
  • Up to 4 body matrices
  • Peripheral Matrix Array, 8-channel knee, shoulder array, wrist array and breast biopsy
  • CP loop (S/L) and head
  • TX/RX head
  • Endorectal
  • Double loop array

Specifications

  • RF Frequency MHz: 123.2
  • Shielding: Active
  • Homogeneity (V-RMS): 0.1
  • Number of Measurement Planes: 24
  • Number of Points Per Plane: 20
  • Field Stability PPM/hr: <0.1
  • Cooling System: Liquid helium
  • Boil-Off Rate l/hr: <0.12
  • Helium Refill: ~9 months
  • Passive on Installation: Yes
  • Number of Shim Plates: 16 trays, 15 pockets per tray
  • Active with Patient in Position: Yes 3D autoshim
  • Number of Independent Channels: 3 linear terms via gradient offsets with 20 coils, Additional 5 non linear second order terms with 32 coils*
  • Width (Patient Aperture at Narrowest): 60 cm
  • Length (Patient Aperture at Narrowest): 142 cm
  • Height (Couch to Pole) (Patient Aperture at Narrowest): 45.5 cm
  • Length (with Covers) (Patient Aperture at Maximum): 213 cm
  • Min Height: 57 cm
  • Max Height: 100 cm
  • Table Top Width: 54 cm
  • Body Mass Limit: 200 kg (250 kg with no vertical movement)
  • Depth (with Covers) (z): 213 cm
  • Width (with Covers) (x): 244 cm
  • Height (with Covers) (y): 221 cm
  • Radial (x,y) 0.5 mT Fringe Field: 3.4 m
  • Axial (z) 0.5 mT Fringe Field: 5.9 m
  • Minimum Installed Area m2*: <33
  • Minimum Ceiling Height cm*: 273

*To include 0.5 mT fringe field

  • Number of Cabinets: 2 cabinets, additional RF-amplifier
  • Total Width: 156 cm
  • Maximum Depth: 65 cm
  • Maximum Height: 197 cm
  • Cooling System: Water
  • Name/Type/Version of the System: Tim Tim Tim [102×8] [102×18] [102×32]
  • Number of Independent RF Receiver Channels: 8 18 32
  • Bandwidth of Each Independent RF Receiver Channel (MHz): 1 1 1
  • Number of Analog-to-Digital Converters for Each Independent RF Channel: 1 1 1
  • Sampling Frequency of Each Analog-to-Digital Converter (MHz): 10 10 10
  • Shielding: Active
  • Single Axis Maximum Amplitude mT/m – x (Horizontal): 40
  • Single Axis Maximum Amplitude mT/m – y (Vertical): 40
  • Single Axis Maximum Amplitude mT/m – z (Along the Bore Axis): 45
  • Single Axis Slew Rate mT/m/ms – x (Horizontal): 200
  • Single Axis Slew Rate mT/m/ms – y (Vertical): 200
  • Single Axis Slew Rate mT/m/ms – z (Along the Bore Axis): 200
  • Duty Cycle at Max Amplitude %: 40; 40;
  • Minimum FOV: 5 mm
  • Maximum Imaging Matrix: 1024 x 1024
  • Minimum 2D Slice Thickness: 0.1 mm
  • Minimum 3D Slice Thickness: 0.05 mm