Navigation Menu

Products by Category

HomeAbout UsProductsServicesRefurbishingContact Us

 
 

 

 


Canon CXDI-40EC Premium General Detector

The Canon CXDI-40EC, a premium DR system offers fast, filmless X-ray capture for a broad range of radiographic applications, including trauma, general, pediatric and orthopedic exams. This new DR system consists of a Canon Amorphous Silicon (a-Si) Flat Panel Detector and a Cesium Iodide (CsI) scintillator, allowing for extremely effective X-ray absorption and high signal-to-noise performance.

The CXDI-40EC is equipped with a large 17-inch x 17-inch imaging area for capturing images large and small without the need to adjust detector orientation. The versatile CXDI-40EC can upgrade your existing radiographic suite or be retrofit to a new upright stand, universal stand or bucky table.


Download Brochure

Features

The high-resolution, premium CXDI-40EC detector features maximum image quality and minimal x-ray exposure to patients with state-of-the-art LANMIT 8 (Large Area New-MIS Sensor and TFT) technology, making the CXDI-40EC perfectly suited for pediatric and orthopedic use.

Large Imaging Area
The large 17” x 17” imaging area flexibly accommodates both large and small format anatomical views without the need to adjust detector orientation.

Enhanced Productivity
The technologist selects the necessary views to complete the requested exam or the requested views are automatically selected through communication between the DICOM worklist and the RIS.

Immediate Results
An on-screen preview image is available just 3-5 seconds after exposure. Sequential images can be acquired in as little as 6 seconds.

Seamless Network Integration
Ethernet 10/100/1000 Base T connectivity with DICOM 3.0 compatibility enables seamless data transfer to any DICOM hard copy output device, PACS, or HIS/RIS for efficient printing, archiving, and remote viewing of Images. The system complies with IHE (Integrating the Healthcare Enterprise) standards for interoperability to ensure easy, effective integration with the hospital network.

Specifications

Detector  
  Detector LANMIT 8, Amorphous Silicon and Scintillator
  Scintillator Cesium Iodide (CsI)
  Pixel Pitch 160 x 160 microns
  Pixels 2,688 x 2,688 (7.2 million)
  Imaging Area 17” x 17” (43 cm x 43 cm)
 
Image Acquisition and Processing  
  A/D 14 bit
  Grayscale 12 bit (4,096 gradations)
  Preview Image 3-5 seconds
  Refresh Cycle 6-10 seconds
 
Data Output and Network Connection  
  Network Ethernet 10/100/1000 Base T, RJ45
  Data Output DICOM 3.0 compliant, Print Management Service Class (SCU), Storage Service Class (SCU), and others
 
Electrical and Environmental  
  Voltage 100V, 120V, 230/240V (50/60Hz)
  Power Consumption 200VA maximum (Detector unit only)
  Operating Environment (Detector unit only)
41-95°F (5-36°C), 30-75% RH (non-condensing)
 
Physical Characteristics  
  Dimensions WxLxH 21.7” x 21.7”x2.7” (550 x 550 x 67.5 mm) (Detector unit only)
  Weight Approximately 44 lbs. (20 kg) without grid

 


 